Hook up light pen to back of scope, turn on power.
Scan: Initialize
Scan: Version 1.1-à3 beeps= newer version than 1.1 installed
Set Home position.
Move stage down till beep is heard
Scan: Home position
Set focus position
Make sure Eyepieces are "zeroed"
Using 10X obj. move stage up till sample is in focus
Scan: Reference Focal Position Setting
Identify objectives
Scan: Data setting address 1
Scan: plan apo 4x (or whatever objective is in position 1.
Repeat this for each position.
Nothing in a position, don't need to do anything
Identify objective turret (Nosepiece)
Scan: motorized
Scan: 7 place
Set Focus offset/parafocalizing
Using 4X obj. focus on sample
Scan: Focal position setting
Repeat with all objectives
Reboot, ie turn power on then off
Set objective selector switch if desired- this allows you to run filter cubes from side switches. Default is that both sides run objective turret.
Kohler Illumination/ Field Diaphragm Adjustment
Adjust condenser up/down till octagonal edges of field diaphragm are in sharp focus.
Center the field diaphragm using adjuster knobs located on condenser, one on each side.
Open field diaphragm (on bottom right side) till diaphragm is just outside field of view |